Abstract

Objective: Reliability of test results is critical in the Real-time PCR test used in the diagnosis of COVID-19. In our study, we aimed to investigate the compatibility of the three most commonly used commercial diagnostic kits in Turkiye in determining the SARS-CoV-2 genome in different Real-time PCR systems and to provide a perspective to laboratory experts in this regard.

Methods: In this study, oro/nasopharyngeal swab samples of 46 COVID-19 patients were used. The samples were analyzed on Rotor-Gene, Bio-Rad and Abbott Real-time PCR devices using Diagnovital, Bio-Speedy and Coronex diagnostic kits to detect SARS-CoV-2 genome. The Ct values were recorded. The kits used were all validated with the devices except with Abbott, only.

Results: The highest and the lowest Ct values were observed in the Abbott–Diagnovital and the Rotor-Gene– Bio-Speedy pairs, respectively. The same result was found in samples with low viral load in the study. When all assessments were evaluated, it was observed that Diagnovital kit gave the highest Ct values among the kits.

Conclusion: For SARS-CoV-2 detection, the pairs of the Rotor-Gene and Bio-Speedy, and the Bio-Rad–Bio-Speedy were found to be more compatible than the other device-kit pairs. False negative results or high Ct values may lead to early discharge of patients or quarantine failures, which may increase the infection. Thus, as the different performances of Diagnovital kit on the three Real-time PCR instruments indicates, when a PCR kit is planned for routine diagnostic use, it is crucial to assess its compatibility with the instrument in the laboratory.

Keywords: Reagent Kits, Diagnostic, PCR, SARS-CoV-2

References

  1. Samudrala PK, Kumar P, Choudhary K, et al. Virology, pathogenesis, diagnosis and in-line treatment of COVID-19. Eur J Pharmacol. 2020;883:173375. https://doi.org/10.1016/j.ejphar.2020.173375
  2. Cucinotta D, Vanelli M. WHO Declares COVID-19 a Pandemic. Acta Biomed. 2020;91(1):157-60. https://doi.org/10.23750/abm.v91i1.9397
  3. World Health Organization (WHO). Historical working definitions and primary actions for SARS-CoV-2 variants. [https://www.who.int/publications/m/item/historical-working-definitions-and-primary-actions-for-sars-cov-2-variants] (Erişim tarihi: 06.09.2023)
  4. Pollard CA, Morran MP, Nestor-Kalinoski AL. The COVID-19 pandemic: a global health crisis. Physiol Genomics. 2020;52(11):549-57. https://doi.org/10.1152/physiolgenomics.00089.2020
  5. World Health Organization (WHO). Recommendations for national SARS-CoV-2 testing strategies and diagnostic capacities: Interim guidance, 25 June 2021. [https://apps.who.int/iris/bitstream/handle/10665/342002/WHO-2019-nCoV-lab-testing-2021.1-eng.pdf?sequence=1&isAllowed=y](Erişimtarihi: 22.08.2023)
  6. Quilty BJ, Clifford S, Hellewell J, et al. Quarantine and testing strategies in contact tracing for SARS- CoV-2: a modelling study. Lancet Public Health. 2021;6(3):e175-83. https://doi.org/10.1016/S2468-2667(20)30308-X
  7. Puylaert CAJ, Scheijmans JCG, Borgstein ABJ, et al. Yield of screening for COVID-19 in asymptomatic patients before elective or emergency surgery using chest CT and RT-PCR (SCOUT): Multicenter study. Ann Surg. 2020;272(6):919-24. https://doi.org/10.1097/SLA.0000000000004218
  8. Harrington WN, Kackos CM, Webby RJ. The evolution and future of influenza pandemic preparedness. Exp Mol Med. 2021;53(5):737-49. https://doi.org/10.1038/s12276-021-00603-0
  9. Eberle U, Wimmer C, Huber I, et al. Comparison of nine different commercially available molecular assays for detection of SARS-CoV-2 RNA. Eur J Clin Microbiol Infect Dis. 2021;40(6):1303-8. https://doi.org/10.1007/s10096-021-04159-9
  10. Rabaan AA, Tirupathi R, Sule AA, et al. Viral dynamics and real-time RT-PCR Ct values correlation with disease severity in COVID-19. Diagnostics (Basel). 2021;11(6):1091. https://doi.org/10.3390/diagnostics11061091
  11. Watson J, Whiting PF, Brush JE. Interpreting a covid-19 test result. BMJ 2020;369:m1808. https://doi.org/10.1136/bmj.m1808
  12. La Scola B, Le Bideau M, Andreani J, et al. Viral RNA load as determined by cell culture as a management tool for discharge of SARS-CoV-2 patients from infectious disease wards. Eur J Clin Microbiol Infect Dis. 2020;39(6):1059-61. https://doi.org/10.1007/s10096-020-03913-9
  13. Mulahuseinovic N, Salihefendic L, Durgut S, Kandic E, Konjhodzic R. Comparative analysis of SARS-CoV-2 detection kits. Acta Inform Med. 2022;30(2):110-4. https://doi.org/10.5455/aim.2022.30.110-114
  14. Rahbari R, Moradi N, Abdi M. rRT-PCR for SARS-CoV-2: Analytical considerations. Clin Chim Acta. 2021;516:1-7. https://doi.org/10.1016/j.cca.2021.01.011

How to cite

1.
Albayrak H, Zeyrek FY, Yiğin A. Real-time PCR detection of SARS-CoV-2: Performances Vary Based on Instrumentation and Kits. Turk Mikrobiyol Cemiy Derg. 2025;55(1):43-49. https://doi.org/10.54453/TMCD.2025.01488