Abstract
Objective: Microbiological screening of blood donors for agents of transfusion transmitted infections is important and necessary for ensuring the transfusion safety. The aim of this study was to prepare internal quality control (IQC) sera for HBsAg screening of blood donors by using the human plasma collected from blood donors with detectable HbsAg levels.
Methods: Quarantined fresh plasmas separated from the whole blood samples of donors with repeatedly reactive screening tests for HbsAg with high S/CO ratio (>1000) and detectable (>50IU/ml) HBV DNA were used for the preparation of IQC sera. In vitro coagulation of pooled plasma was performed by using calcium chloride and bovine thrombin. Following the removal of coagulum, dextrane sulphate was added and precipitation of lipids was carried out in a series of centrifugation processes. Samples from the sera pool were diluted with PBS and diluted samples were tested for HbsAg reactivity. Recorded S/CO ratio of IQC samples from each category was applied to Levey-Jennings control charts and evaluated according to Westgard rules.
Results: In descriptive analysis of high, medium and low reactive IQC samples the mean was found to be 3323.75, 1725.57 and 219.57; the median was determined as 3401.76, 1753.48 and 217.34; the standard deviation was calculated as 271.19, 210.62 and 14.93, respectively. Coeffcients of variation for high, medium and low reactive IQC samples were determined as %8.1, %12.2 and %7.6, respectively.
Conclusion: In concordance with ISO15189 standards that suggest the use of independent run controls for immunoassays, it would be beneficial to prepare IQC samples from collected blood in blood banks.
Keywords: HbsAg, Blood Donor, Internal Quality Control
References
- Clark RB, Lewinski MA, Loeffelholz MJ, Tibbetts RJ. Cumitech 31A. Verification and validation of procedures in the clinical microbiology laboratory. ASM Press, Washington, DC. 2009.
- Public Health England. Quality assurance in the diagnostic virology and serology laboratory. Standards Unit Microbiology Services. 2015.
- Sepulveda JL. Variation, errors, and quality in the clinical laboratory. In: Dasgupta A, Sepulveda JL (Eds.) Accurate Results in the Clinical Laboratory. (2nd Ed). Elsevier 2019:3-9.
- Proksch GJ, Bonderman DP. Preparation of optically clear lyophilized human serum for use in preparing control material. Clin Chem. 1976;22(4):456-60.
- Fiedler SA, Oberle D, Chudy M, et al. Effectiveness of blood donor screening by HIV, HCV, HBV‐NAT assays, as well as HBsAg and anti‐HBc immunoassays in Germany (2008–2015). Vox Sang. 2019;114(5):443- 50. https://doi.org/10.1111/vox.12770
- Dodd RY, Nguyen ML, Krysztof DE, Notari EP, Stramer SL. Blood donor testing for hepatitis B virus in the United States: is there a case for continuation of hepatitis B surface antigen detection? Transfusion. 2018;58(9):2166-70. https://doi.org/10.1111/trf.14784
- Örüç NE, Yenicesu İ. Ulusal Kan ve Kan Bileşenleri Hazırlama, Kullanım ve Kalite Güvencesi Rehberi. 2016. Avrupa Birliği’nin IPA-I finansal desteği ile Kan Tedarik Sisteminin Güçlendirilmesi Teknik Destek Projesi. https://sbu.saglik.gov.tr/Ekutuphane/Yayin/523 (Erişim: Aralık 2020).
- Allain JP, Cox L. Challenges in hepatitis B detection among blood donors. Curr Opin Hematol. 2011;18(6):461-6. https://doi.org/10.1097/MOH.0b013e32834bac10
- Candotti D, Laperche S. Hepatitis B virus blood screening: Need for reappraisal of blood safety measures?. Front Med. 2018;5:29. https://doi.org/10.3389/fmed.2018.00029
- Kuhns MC, Kleinman SH, McNamara AL, et al. Lack of correlation between HBsAg and HBV DNA levels in blood donors who test positive for HBsAg and anti- HBc: implications for future HBV screening policy. Transfusion. 2004;44(9):1332-9. https://doi.org/10.1111/j.1537-2995.2004.04055.x
- Pronier C, Candotti D, Boizeau L, Bomo J, Laperche S, Thibault V. The contribution of more sensitive hepatitis B surface antigen assays to detecting and monitoring hepatitis B infection. J Clin Virol. 2020;129:104507. https://doi.org/10.1016/j.jcv.2020.104507
- Amini A, Varsaneux O, Kelly H, et al. Diagnostic accuracy of tests to detect hepatitis B surface antigen: a systematic review of the literature and meta-analysis. BMC Infect Dis. 2017;17(Suppl 1):698. https://doi.org/10.1186/s12879-017-2772-3
- Wilkinson DE, Seiz PL, Schüttler CG, et al. International collaborative study on the 3rd WHO International Standard for hepatitis B surface antigen. J Clin Virol. 2016;82:173-80. https://doi.org/10.1016/j.jcv.2016.06.003
- Liu TW, Yeh ML, Huang CF, et al. Clinical performance of a new hepatitis B surface antigen quantitative assay with automatic dilution. Kaohsiung J Med Sci. 2015;31(1):26-33. https://doi.org/10.1016/j.kjms.2014.10.007
- Wang L, Chen W, Yu Y. The performance of the Abbott i2000 for measuring serum markers of infectious diseases. J Clin Lab Anal. 2017;31(1):e22015. https://doi.org/10.1002/jcla.22015
- Kupek E. Residual risk of hepatitis-B-infected blood donations: estimation methods and perspectives. Int Sch Res Notices. 2013;2013:839896. https://doi.org/10.5402/2013/839896
- Busch MP. Should HBV DNA NAT replace HBsAg and/ or anti-HBc screening of blood donors?. Transfus Clin Biol. 2004;11(1):26-32. https://doi.org/10.1016/j.tracli.2003.12.003
- Gerlich WH, Glebe D, Schüttler CG. Hepatitis B viral safety of blood donations: new gaps identified. Ann Blood. 2018;3:38. https://doi.org/10.21037/aob.2018.09.03
- Godbey EA, Thibodeaux SR. Ensuring safety of the blood supply in the United States: Donor screening, testing, emerging pathogens, and pathogen inactivation. Semin Hematol. 2019;56(4):229-35. https://doi.org/10.1053/j.seminhematol.2019.11.004
- Wang Z, Zeng J, Li T, et al. Prevalence of hepatitis B surface antigen (HBsAg) in a blood donor population born prior to and after implementation of universal HBV vaccination in Shenzhen, China. BMC Infect Dis. 2016;16:498. https://doi.org/10.1186/s12879-016-1834-2
- Candotti D, Assennato SM, Laperche S, Allain JP, Levicnik-Stezinar S. Multiple HBV transfusion transmissions from undetected occult infections: revising the minimal infectious dose. Gut. 2019;68(2):313-21. https://doi.org/10.1136/gutjnl-2018-316490
Copyright and License
Copyright © 2022 Turkish Society of Microbiology. This is an open access article distributed under the terms of the Creative Commons Attribution 4.0 International License (CC BY 4.0), which permits unrestricted use, distribution, and reproduction in any medium or format, provided the original work is properly cited.