Abstract

Objective: Tuberculosis has high morbidity and mortality, so it requires rapid diagnosis and treatment. Microscopic examination has an important place in the diagnosis of Mycobacterium tuberculosis infections. The aim of this study is to compare two different protocols of an automated mycobacteria staining system with the manual staining method in terms of performance and quality, and to show that automated system can be used without cross-contamination.

Method: In this study, acidoresistance staining positive (ARS+) 100 pulmonary specimens which were sent to a university hospital laboratory for routine analysis were examined. Clinical samples were decontaminated, then a total of three smears were prepared from these samples.These smears were stained according to manual fluorescence staining protocol, conventional automated staining protocol and short protocol respectively. Smears were examined in terms of the presence of artefacts, the color brightness and the ease of evaluation.

Results: Positive controls evaluated in the study were positive,and negative controls were negative, any crosscontamination was not detected in any sample in the automated staining system. The agreement between the results obtained in the manual staining protocol and the automated staining system (conventional and short protocol) was 84% and 82% for smear grade, 92% and 85% for presence of artefacts, 86% and 89% for brightness of the color, 91% and 89% for the ease of evaluation, respectively. The Cohens kappa coefficients between the manual protocol, the automated conventional and short protocols were 0.79 and 0.76 that are considered as indicative of excellent agreement, respectively.

Conclusion: The automated staining system can be used because it decreases the chemical contamination, reduces workload in specialized laboratories which perform medium to high numbers of staining in their daily practice and improves standardization of staining methods.

Keywords: Fluorescent dyes, microscopy, tuberculosis

References

  1. World Health Organization. Global tuberculosis report 2018. https://www.who.int/tb/publications/global_report/en/ (Erişim tarihi: Aralık 2018)
  2. Mugusi F, Villamor E, Urassa W, Saathoff E, Bosch RJ, Fawzi WW. HIV co-infection, CD4 cell counts and clinical correlates of bacillary density in pulmonary tuberculosis. Int J Tuberc Lung Dis. 2006;10(6):663-9.
  3. Chandra TJ, Selvaraj R, Sharma YV. Same day sputum smear microscopy for the diagnosis of pulmonary tuberculosis: Ziehl-Neelsen versus fluorescent staining. J Family Med Prim Care. 2015;4(4):525-8. https://doi.org/10.4103/2249-4863.174273
  4. Tiemersma EW, van der Werf MJ, Borgdorff MW, Williams BG, Nagelkerke NJ. Natural history of tuberculosis: duration and fatality of untreated pulmonary tuberculosis in HIV negative patients: a systematic review. PloS One. 2011;6(4):e17601. https://doi.org/10.1371/journal.pone.0017601
  5. RAL STAINER-clinical diagnostics products. https://www.biomerieux-diagnostics.com/sites/clinic/files/9304423_002_gb_a_ral_stainer_ok.pdf (Erişim Tarihi: Aralık 2018).
  6. Kent PT, Kubica GP. Public Health Mycobacteriology: A guide for the Level III Laboratory PB86216546. Centers for Disease Control, Atlanta, GA, 1985.
  7. Landis JR, Koch GG. The measurement of observer agreement for categorical data. Biometrics. 1977;33(1):159-74. https://doi.org/10.2307/2529310
  8. Akobeng AK. Understanding randomised controlled trials. Arch Dis Child. 2005;90(8):840-4. https://doi.org/10.1136/adc.2004.058222
  9. Gérôme P, Fabre M, Soler C. Evaluation of the Fluo-RAL (RAL) kit for the identification of mycobacteria by fluorescence microscopy. Pathol Biol (Paris). 2011;59(5): 245-7. https://doi.org/10.1016/j.patbio.2009.10.012
  10. Getahun H, Harrington M, O’Brien R, Nunn P. Diagnosis of smear-negative pulmonary tuberculosis in people with HIV infection or AIDS in resource-constrained settings: informing urgent policy changes. Lancet. 2007;369(9578):2042-9. https://doi.org/10.1016/S0140-6736(07)60284-0

How to cite

1.
Çavuşoğlu C, Arslan A. Comparison of Manual (TB Fluorescent Stain Kit) and Automated (RAL Stainer) Fluorescent Staining Methods for Direct Microscopic Examination of Respiratory Tract Samples of Patients with Suspected Mycobacterial Infection. Turk Mikrobiyol Cemiy Derg. 2019;49(3):118-124. https://doi.org/10.5222/TMCD.2019.118