Abstract

Objective: It was aimed to make a fast screening algorithm by comparing the results obtained by the flow cytometry method with the urine culture results so as to to eliminate cultivation in samples containing a number of leukocytes and bacteria below a threshold value. In the second step of the study; it was evaluated to what extent unnecessary antibiotic usage could be avoided by finalizing the results of negative-contaminated urine specimens in a short-term.

Material and Methods: The urine samples of 995 patients from all clinics submitted to our laboratory for culture were also tested in a flow cytometry device. The results obtained from the flow cytometry device were compared with the culture results. In the second step, the use of antibiotics in 208 patients referred from the urology clinic were recorded, in order to evaluate the empirical use of antibiotics, which were initiated without the culture results were obtained.

Results: The results we obtained with the flow cytometry method showed that there was no need to carry out urine culture in 31% of our samples that were sent from all the clinics. As to the samples obtained from the urology clinic, it was detected that there was no need to carry out urine culture in 29.3%; and 23% of the patients whose urine cultures were detected as no growth on flow cytometry predicted had been started on unnecessary antibiotic therapy.

Conclusion: When this technique is used as a rapid screening test, it will reduce the workload and cost in laboratories in which urine specimens are frequently sent for analysis; also contribute to a certain extent in reducing unnecessary antibiotic use by providing negative results within minutes.

Keywords: Comparison: flow cytometry and urine cultures

References

  1. Okada H, Sakai Y, Miyazaki S, Arakawa S, Hamaguchi Y, Kamidono S. Detection of significant bacteriuria by automated urinalysis using flow cytometry. J Clin Microbiol. 2000;38(8):2870-2.
  2. Tille PM. Infections of the urinary tract. In. Bailey & Scott’s Diagnostic Microbiology, 13th edn. Elsevier, St. Louis, 2014:919-30.
  3. Sterry-Blunt RE, Randall KS, Doughton MJ, Aliyu SH, Enoch DA. Screening urine samples for the absence of urinary tract infection using the sediMAX automated microscopy analyser. J Med Microbiol. 2015;64(6):605-9. https://doi.org/10.1099/jmm.0.000064
  4. Kadkhoda K, Manickam K, Degagne P, et al. UF-1000i flow cytometry is an effective screening method for urine specimens. Diagn Microbiol Infect Dis. 2011;69(2):130-6. https://doi.org/10.1016/j.diagmicrobio.2010.09.013
  5. Simerville JA, Maxted WC, Pahira JJ. Urinalysis: a comprehensive review. Am Fam Physician. 2005;71(6): 1153-62.
  6. Boonen KJ, Koldewijn EL, Arents NL, Raaymakers PA, Scharnhorst V. Urine flow cytometry as a primary screening method to exclude urinary tract infections. World J Urol. 2013;31(3):547-51. https://doi.org/10.1007/s00345-012-0883-4
  7. Manoni F, Fornasiero L, Ercolin M, et al. Cutoff values for bacteria and leukocytes for urine flow cytometer Sysmex UF-1000i in urinary tract infections. Diagn Microbiol Infect Dis. 2009;65(2):103-7. https://doi.org/10.1016/j.diagmicrobio.2009.06.003
  8. Jolkkonen S, Paattiniemi EL, Kärpänoja P, Sarkkinen H. Screening of urine samples by flow cytometry redu-ces the need for culture. J Clin Microbiol. 2010;48(9):3117-21. https://doi.org/10.1128/JCM.00617-10
  9. Manoni F, Valverde S, Antico F, Salvadego MM, Giacomini A, Gessoni G. Field evaluation of a second-generation cytometer UF-100 in diagnosis of acute urinary tract infections in adult patients. Clin Microbiol Infect. 2002;8(10):662-8. https://doi.org/10.1046/j.1469-0691.2002.00452.x
  10. Zaman Z, Roggeman S, Verhaegen J. Unsatisfactory performance of flow cytometer UF-100 and urine strips in predicting outcome of urine cultures. J Clin Microbiol. 2001;39(11):4169-71. https://doi.org/10.1128/JCM.39.11.4169-4171.2001
  11. Giesen CD, Greeno AM, Thompson KA, Patel R, Jenkins SM, Lieske JC. Performance of flow cytometry to screen urine for bacteria and white blood cells prior to urine culture. Clin Biochem. 2013;46(9):810-3. https://doi.org/10.1016/j.clinbiochem.2013.03.005
  12. Ilki A, Ayas R, Ozsoy S, Soyletir G. Cost-effectiveness of a new system in ruling out negative urine cultures on the day of administration. Eur J Clin Microbiol Infect Dis. 2017;36(7):1119-23. https://doi.org/10.1007/s10096-017-2898-7
  13. Pieretti B, Brunati P, Pini B, et al. Diagnosis of bacteri-uria and leukocyturia by automated flow cytometry compared with urine culture. J Clin Microbiol. 2010;48(11):3990-6. https://doi.org/10.1128/JCM.00975-10
  14. Broeren MA, Bahçeci S, Vader HL, Arents NL. Screening for urinary tract infection with the Sysmex UF-1000i urine flow cytometer. J Clin Microbiol. 2011;49(3):1025-9. https://doi.org/10.1128/JCM.01669-10
  15. Wang J, Zhang Y, Xu D, Shao W, Lu Y. Evaluation of the Sysmex UF-1000i for the diagnosis of urinary tract infection. Am J Clin Pathol. 2010;133(4):577-82. https://doi.org/10.1309/AJCP1GT2JXOCQBCZ
  16. Costelloe C, Metcalfe C, Lovering A, Mant D, Hay AD. Effect of antibiotic prescribing in primary care on antimicrobial resistance in individual patients: syste-matic review and meta-analysis. BMJ. 2010;340: c2096. https://doi.org/10.1136/bmj.c2096
  17. Thompson R, Gammie A, Lewis D, Smith R, Edwards C. Evidence review: automated urine screening systems: CEP10030. NHS Purchas Supply Agency Informing Procurement. 2010:1-46.
  18. Bartlet R, Galen R. Predictive value of urine culture. J Infect Dis. 1983;79(6):179-82. https://doi.org/10.1093/ajcp/79.6.756

How to cite

1.
Üzmez E, Yağcı S, Yücel M, Keseroğlu BB, Karakoç AE, Dinç B. Comparison of Results of Urine Cultures and a Flow Cytometric Device That Analyses Urinary Leukocytes/ Bacteria. Turk Mikrobiyol Cemiy Derg. 2018;48(1):78-85. https://doi.org/10.5222/TMCD.2018.078